Fluorescence-based Oxygen Detector Transmitter E2608-O2-L

Fluorescence-based oxygen detector-transmitter is used to monitor the absence of oxygen in gas mixtures for specific purposes as well as to avoid a crucial drop in O2 concentration.

Fluorescence-based Oxygen Detector-Transmitter model VAC-E2608-O2-L

  1. VAC-E2608-O2-L is a long-lasting optical sensor.
  2. Range of measurement for this fluorescence-based oxygen detector-transmitter is 0-25% vol O2.
  3. Resolution of measurement of VAC-E2608-O2-L is 0.01% vol O2.
  4. Default alarm set-points is 19.5 or 18.0% vol O2.
  5. It has 2 analogue outputs settable at 4-20 mA or 0-10 V with RS485 Modbus RTU digital interface.

Specifications of Fluorescence-based oxygen detector-transmitter model VAC-E2608-O2-L:

  1. Calibration for VAC-E2608-O2-L is Oxygen O2.
  2. Sensing method used is Optical (fluorescence-based).
  3. The fluorescence-based oxygen detector-transmitter model VAC-E2608-O2-L sampling method is diffusion.
  4. Detection range for this fluorescence-based oxygen detector-transmitter is 0-25% vol O2.
  5. Resolution range of VAC-E2608-O2-L is 0.01% vol O2.
  6. It has an accuracy of less than 2% FS.
  7. The signal is updated every 1 second.
  8. Response time is less than 30 seconds.
  9. VAC-E2608-O2-L is a maintenance free sensor.
  10. VAC-E2608-O2-L sensor lifetime is more than 5 years.
  11. Self-diagnostics with full functionality check at start-up.
  12. Warm-up time for VAC-E2608-O2-L is less than or equal to 1 minute.
  13. The power supply of this fluorescence-based oxygen detector-transmitter model VAC-E2608-O2-L is 11-30 VDC, 24 VAC or 90-265 VAC.
  14. The power consumption is less than 2 VA.
  15. VAC-E2608-O2-L has RS485 Modbus RTU digital interface.
  16. It has 2 analogue outputs settable to 4-20 mA or 0-10 V.
  17. The recommended output scale width is 20-100% of the range with more than 10 × resolution in any case.
  18. It has 2 SPST output relays (closing contact) with 250 VAC or 30 VDC at 5 A max current.
  19. The default warning set-points of this fluorescence-based oxygen detector-transmitter model VAC-E2608-O2-L at RE1 (High) is set at 19.5%, release 20.5%.
  20. The default warning set-points of this fluorescence-based oxygen detector-transmitter model VAC-E2608-O2-L at RE2 (Low) is set at 18%, release 19%.
  21. It has a light grey, wall-mounted ABS plastic with an IP65 safety rating.
  22. The dimensions of this fluorescence-based oxygen detector-transmitter model VAC-E2608-O2-L is H85 × W82 × D55 millimetres.
  23. It has an IP65 secured remote sensor probe, shielded cable. The cable’s default length is 3.0 meters.
  24. It can be operated at industrial indoor and outdoor locations.
  25. Operating conditions of this fluorescence-based oxygen detector-transmitter model VAC-E2608-O2-L is -30 to 60 degree Celsius at 15-90% relative humidity, with 0.9-1.1 atmospheric pressure, non-ATEX rated areas, non-aggressive atmosphere.

*if no thermal fluctuations occur

We supply fluorescence-based oxygen detector-transmitter across the world covering Europe, USA, Latin America, Asia, Australia, New Zealand, Canada etc.

In the Middle East, we supply fluorescence-based oxygen detector-transmitter in the United Arab Emirates, Kingdom of Saudi Arabia, Oman, Qatar, Bahrain, Iraq, Lebanon, Egypt, Jordan, Libya, Tunisia etc.

In Africa, we supply fluorescence-based oxygen detector-transmitter in countries including Kenya, Algeria, Tanzania, Djibouti, Chad, Ghana, Rwanda, Uganda, Nigeria, Ethiopia, Morocco, Angola, South Africa etc.

Also, we supply fluorescence-based oxygen detector-transmitter to India, Bangladesh, Pakistan, Afghanistan, Kazakhstan, Srilanka, Maldives and Cyprus.